<原著>腰椎骨密度測定法での患者被曝線量 : 二重エネルギーX線吸収法(QDR-4500)と腰椎一般撮影の比較
URI | http://harp.lib.hiroshima-u.ac.jp/pu-hiroshima/metadata/3545 | ||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
ファイル |
KJ00000044640.pdf
( 327.0 KB )
公開日
:2009-02-12
|
||||||||||||||||||||||||||||||||||||
タイトル |
<原著>腰椎骨密度測定法での患者被曝線量 : 二重エネルギーX線吸収法(QDR-4500)と腰椎一般撮影の比較
|
||||||||||||||||||||||||||||||||||||
別タイトル |
<ORIGINAL ARTICLES>Comparison of the amount of absorbed dose using the DEXA for lumbar spine bone mineral density measurement
|
||||||||||||||||||||||||||||||||||||
著者 |
|
||||||||||||||||||||||||||||||||||||
キーワード |
二重エネルギーX線吸収法(DEXA)
被曝線量
臓器
TLD
骨密度(BMD)
dual-energy X-ray absorptiometry (DEXA)
absorbed dose
organ
bone mineral density (BMD)
|
||||||||||||||||||||||||||||||||||||
抄録 |
骨密度測定から受ける患者の被曝線量を人体ファントムを使った方法で測定を行った。今回は腰椎骨密度測定で行われる2種類の方法(DEXA法, 腰椎一般撮影)で熱ルミネセンス線量計(TLD)を用いて, 臓器の吸収線量を測定し, 以前の研究から導かれた近似式と比較検討した。腰椎骨密度測定の照射野内において, 臓器の線量は実測値と計算値との差が10%未満の誤差であった。これは, 個々のTLDのばらつき±20%以内から考えて大きな誤差ではないと考えられる。骨密度測定方法および装置から受ける臓器の被曝線量は計算によっても求めることが出来るといえる。計算値の比較では, X線が背側から入射する装置よりも腹側から入射する装置および検査法の方が臓器の吸収線量が大きかった。その結果, 骨密度測定は骨折の危険性の高い閉経後の女性への適応があると考えられる。しかし, 若年者の骨密度のスクリーニング検査としては, 注意深く考えなければならない。骨密度測定方法の利点と被曝線量を被検者の年齢を加味した上で, どの方法および装置を用いるか選択する必要がある。 Patient dose in bone mineral density examinations was measured using a body phantom. In this study, the organ absorbed dose from two different methods of lumbar bone mineral densitometory, dual-energy x-ray absorptiometry (DEXA) and lumbar vertebral radiography, was measured using a thermo-luminescence dosimeter (TLD). Measured values were compared with the approximations derived from our previous study. Differences between the measured and calculated values was less than 10% in the estimated field area. This error seems adequately small, considering that fluctuation of measured values with TLD is within ± 20%. Consequently, the organ absorbed dose can be obtained with calculation, even if the method or device is different. In the comparison of calculated values, the organ absorbed dose for the method or device in which x-rays enter the human body in the dorso-ventral projection was greater than in the ventro-dorsal projection. Bone mineral density measurement is generally applied to females after menopause, who are in risk of bone fracture. However, application to young generation screening should be considered carefully. The method and the device must be carefully chosen, considering each merit, patient dose and patient age. |
||||||||||||||||||||||||||||||||||||
内容記述 |
国立情報学研究所で電子化 |
||||||||||||||||||||||||||||||||||||
掲載雑誌名 |
広島県立保健福祉短期大学紀要
|
||||||||||||||||||||||||||||||||||||
巻 |
5
|
||||||||||||||||||||||||||||||||||||
号 |
1
|
||||||||||||||||||||||||||||||||||||
開始ページ |
43
|
||||||||||||||||||||||||||||||||||||
終了ページ |
48
|
||||||||||||||||||||||||||||||||||||
出版年月日 |
2000-03
|
||||||||||||||||||||||||||||||||||||
出版者 |
広島県立保健福祉短期大学
|
||||||||||||||||||||||||||||||||||||
ISSN |
1342-0070
|
||||||||||||||||||||||||||||||||||||
NCID |
AA11318754
|
||||||||||||||||||||||||||||||||||||
本文言語 |
日本語
|
||||||||||||||||||||||||||||||||||||
資料タイプ |
紀要論文
|
||||||||||||||||||||||||||||||||||||
著者版フラグ |
出版社版
|
||||||||||||||||||||||||||||||||||||
ソートキー |
08
|
||||||||||||||||||||||||||||||||||||
旧URI | |||||||||||||||||||||||||||||||||||||
区分 |
pu-hiroshima
|